Bid Publish Date
20-Aug-2026, 1:32 pm
Bid End Date
18-Sep-2026, 10:00 am
EMD
₹44,550
Location
Progress
Public procurement opportunity for Eastern Railway ICU Ventilators (V2) (Q2) in KOLKATA, WEST BENGAL. Quantity: 1 issued by. Submission Deadline: 18-09-2026 10: 00: 00. View full details and respond.
Quantity
1
Bid Type
Single Packet Bid
Bid Validity
90 (Days)
Bid Type
Service
Evaluation
Total value wise evaluation
Inspection Required
No
Tech Clarification Time
3 Days
EMD Required
Yes
MII Purchase Preference
No
MSE Purchase Preference
Yes
MSE Preference Band
L1+15%
MSE Exemption/Relaxation
No
Startup Exemption/Relaxation
No
Bid Splitting Applied
No
Warranty Period
15 years
Experience Required
3 Year (s)
Arbitration Clause
No
Mediation Clause
No
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
No
Item Category
ICU Ventilators (V2) (Q2)
Advisory Bank
State Bank of India
ePBG Percentage
5%
ePBG Duration (Months)
32
Max Delivery Days
45
Delivery Locations
1
Delivery Cities
Kolkata
Delivery Pincodes
700014
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| JAYEETA BANDYOPADHYAY DUTTA | 700014,MEDICAL DIRECTOR, B R SINGH HOSPITAL EASTERN RAILWAY,MEDICAL DEPARTMENT,SEALDAH | Kolkata | West Bengal | 700014 | 1 | 45 | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
OTHER
CATALOG Specification
OTHER
GEM_GENERAL_TERMS_AND_CONDITIONS
Extended Deadline
18-Sep-2026, 4:30 am
Opening Date
19-Sep-2026, 4:30 am
Discover companies most likely to bid on this tender
Experience Criteria
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
| Category | Specification | Requirement |
|---|---|---|
| PRODUCT INFORMATION | Patient Type | Adult & pediatric |
| PRODUCT INFORMATION | Type of technology | Turbine |
| PRODUCT INFORMATION | Tidal volume in ml | (20 ml or less) to (1500 ml or more) |
| PRODUCT INFORMATION | Respiration rate, breaths/min | up to 80 or more |
| PRODUCT INFORMATION | Trigger mechanism | Pressure /flow |
| PRODUCT INFORMATION | Inspiratory flow rate, L/min | upto 200 or more |
| PRODUCT INFORMATION | Inspiratory pressure, cm H2O | upto 80 or more |
| PRODUCT INFORMATION | IE ratio | 1:10 to 4:1 |
| PRODUCT INFORMATION | Sigh breath function | Yes |
| PRODUCT INFORMATION | Control panel lock | Yes |
| PRODUCT INFORMATION | Facility for double lung ventilation | Yes |
| PATIENT ASSESSMENT TOOLS | Number of loops | 2 loops (P-V, F-V) with facility of saving of 1 Loops for reference |
| PATIENT ASSESSMENT TOOLS | Lung recruitment tools (PV loops) / OLT (Open lung tools) | Yes |
| PATIENT ASSESSMENT TOOLS | Lung mechanics visualization tool | Yes |
| PATIENT ASSESSMENT TOOLS | Capnography/CO2 monitoring | Yes |
| PATIENT ASSESSMENT TOOLS | Esophageal / transpulmonary pressure monitoring | No |
| PATIENT ASSESSMENT TOOLS | APRV(Airway Pressure Release Ventilation) /Bivent/Bi level | Yes |
| PATIENT ASSESSMENT TOOLS | MMV(Mandatory Minute Ventilation) + PSV (Pressure Support Ventilation) /ASV( Adaptive Support Ventilation ) or ALPV(Adaptive Lung Protection Ventilation) or Automode | Yes |
| PATIENT ASSESSMENT TOOLS | PRVC(Pressure-Regulated Volume Control ventilation)/Auto flow | Yes |
| PATIENT ASSESSMENT TOOLS | High flow oxygen therapy | No |
| PATIENT ASSESSMENT TOOLS | Automatic weaning system provided | No |
| PATIENT ASSESSMENT TOOLS | Volume Support | No |
| PATIENT ASSESSMENT TOOLS | RSBI (Rapid Shallow Breathing Index) | Yes |
| PATIENT ASSESSMENT TOOLS | Expiratory Time constant ( Tcexp) | No |
| PATIENT ASSESSMENT TOOLS | Compliance static and/or dynamic | Yes |
| PATIENT ASSESSMENT TOOLS | Inspired and/or expired resistance | Yes |
| PATIENT ASSESSMENT TOOLS | Occlusion pressure | No |
| PATIENT ASSESSMENT TOOLS | Inspiratory and expiratory hold | Yes |
| PATIENT ASSESSMENT TOOLS | Spontaneous frequency | Yes |
| PATIENT ASSESSMENT TOOLS | Total peep, intrinsic peep, extrinsic peep | Yes |
| PATIENT ASSESSMENT TOOLS | Expiratory block is autoclavable and no routine calibration required | Yes |
| PATIENT ASSESSMENT TOOLS | Auto compensation for ET Tube | No |
| PATIENT ASSESSMENT TOOLS | Facility for automatic compliance & Leakage compensation for circuit available | Yes |
| PATIENT ASSESSMENT TOOLS | RS 232 out put port | Yes |
| PATIENT ASSESSMENT TOOLS | USB/Memory card | No |
| PATIENT ASSESSMENT TOOLS | Availability of Remote alarm/display port | Yes |
| PATIENT ASSESSMENT TOOLS | Facility to Report (vent alarms and patient status) | Yes |
| PATIENT ASSESSMENT TOOLS | HL/7 Compliant out put | Yes |
| PATIENT ASSESSMENT TOOLS | Type of coloured display | LED/LCD/TFT |
| PATIENT ASSESSMENT TOOLS | Display should be touch screen | Yes |
| PATIENT ASSESSMENT TOOLS | NIV (Non Invasive Ventilation) to be possible in all modes of ventilation available | Yes |
| PATIENT ASSESSMENT TOOLS | Minimum Size of display (±2%) | 12.0 |
| PATIENT ASSESSMENT TOOLS | Graphic display have automatic scaling facility for waves | Yes |
| HARDWARE AND ELECTRICAL REQUIREMENTS | Provision of UPS | No |
| HARDWARE AND ELECTRICAL REQUIREMENTS | Backup time in hrs | NA if not provided |
| HARDWARE AND ELECTRICAL REQUIREMENTS | Built in air source | Turbine (In built) |
| HARDWARE AND ELECTRICAL REQUIREMENTS | Availability of stand alone compressor | No |
| HARDWARE AND ELECTRICAL REQUIREMENTS | Compressor shall provide an oil free Medical air, with Peak output flow should be minimum 150 LPM and Air quality complying with ISO compressed air purity class | NA (for turbine) |
| HARDWARE AND ELECTRICAL REQUIREMENTS | Seller shall ensure compatibility of compressor with ventilator | NA (for Turbine) |
| HARDWARE AND ELECTRICAL REQUIREMENTS | Nebulizer with particle size less than 3 micron | Yes |
| ADDITIONAL ACCESSORIES | Humidifier: Servo controlled with digital monitoring of inspired gas temperature with heating wire | NA |
| WARRANTY | Warranty (years) | 5 Or higher |