All India Institute Of Medical Sciences (aiims) announces a tender for Multipara Monitor - Low End (V2) (Q2) in PATNA, BIHAR. Quantity: 4. Submission Deadline: 17-09-2026 13: 00: 00. Last date to apply is approaching fast!
Quantity
4
Bid Type
Two Packet Bid
Bid Validity
180 (Days)
Bid Type
Service
Evaluation
Total value wise evaluation
Inspection Required
No
RA Qualification Rule
H1-Highest Priced Bid Elimination
Tech Clarification Time
2 Days
EMD Required
No
MII Purchase Preference
No
MSE Purchase Preference
Yes
MSE Preference Band
L1+15%
MSE Exemption/Relaxation
Yes
Startup Exemption/Relaxation
Yes
Bid Splitting Applied
No
Min. Avg. Annual Turnover
2
OEM Avg. Turnover
10
Past Performance
50 %
Warranty Period
3 days
Experience Required
3 Year (s)
Arbitration Clause
No
Mediation Clause
No
Tender Category
Goods
Bid To RA
No
Bid To RA Enabled
Yes
Item Category
Multipara Monitor - Low End (V2) (Q2)
Advisory Bank
Bank of India
ePBG Percentage
5%
ePBG Duration (Months)
26
Max Delivery Days
30
Delivery Locations
1
Delivery Cities
Patna
Delivery Pincodes
801507
| Consignee | Address | City | State | Pincode | Quantity | Delivery Days | Additional Requirement |
|---|---|---|---|---|---|---|---|
| Neeraj Dwivedi | 801507,AIIMS, Phulwarisarif, PATNA PIN-851507 | Patna | Bihar | 801507 | 4 | 30 | - |
End-to-end support — bid preparation, GeM registration, document filing & compliance by industry experts.
Free consultation · 24h response
Main Document
CATALOG Specification
GEM_GENERAL_TERMS_AND_CONDITIONS
Extended Deadline
17-Sep-2026, 7:30 am
Opening Date
18-Sep-2026, 7:30 am
Discover companies most likely to bid on this tender
Experience Criteria
Past Performance
Bidder Turnover
Certificate (Requested in ATC)
OEM Authorization Certificate
OEM Annual Turnover
Additional Doc 1 (Requested in ATC)
Compliance of BoQ specification and supporting document *In case any bidder is seeking exemption from Experience / Turnover Criteria
the supporting documents to prove his eligibility for exemption must be uploaded for evaluation by the buyer
| Category | Specification | Requirement |
|---|---|---|
| PRODUCT INFORMATION | Type of Monitor | Preconfigured |
| PRODUCT INFORMATION | Monitoring Parameters | ECG, HR, NIBP, PR, Respiratory rate, SpO2, Body temperature |
| PRODUCT INFORMATION | Noise level ( in dB ) | <=50 |
| ECG PARAMETERS | Number of leads of ECG | 5 |
| ECG PARAMETERS | 12-lead interpretation for ECG | No |
| ECG PARAMETERS | Arrhythmia detection and analysis | Yes |
| ECG PARAMETERS | Number of leads analyzed for ECG in single time | 7 |
| ECG PARAMETERS | Heart rate measurement range (atleast) | 30 - 250 BPM |
| ECG PARAMETERS | Accuracy | ± 3 BPM |
| ECG PARAMETERS | Minimum gradation | 1 BPM |
| RESPIRATION PARAMETERS | Respiration Method | Impedance Pneumography |
| RESPIRATION PARAMETERS | Display waveform of respiration | Yes |
| RESPIRATION PARAMETERS | Threshold control for respiration | Yes |
| RESPIRATION PARAMETERS | Respiration rate measurement range (atleast) | 0-100 BPM |
| RESPIRATION PARAMETERS | Minimum gradation of respiratory measurement | 1 BPM |
| PULSE OXIMETRY PARAMETERS | Measurement technique (Declaration of the measurement technique by the seller at the time of bid submission / supply) | Low Perfusion Technology with Motion artifact detection |
| PULSE OXIMETRY PARAMETERS | Probe type provided | Adult |
| PULSE OXIMETRY PARAMETERS | Probe usage | Disposable |
| PULSE OXIMETRY PARAMETERS | SPO2 measurement range | 1 - 100% |
| PULSE OXIMETRY PARAMETERS | Accuracy of SPO2 measurement | ± 2% or better |
| PULSE OXIMETRY PARAMETERS | Minimum Gradation | 1% |
| NIBP PARAMETERS | Measurement technique for NIBP | Step-deflation Oscillometric / Linear Inflation |
| NIBP PARAMETERS | Blood pressure monitoring range | 30 - 300 mm/Hg |
| NIBP PARAMETERS | Gradation | 1 mm/Hg |
| TEMPERATURE PARAMETERS | Number of channels for temperature | 1 |
| TEMPERATURE PARAMETERS | Probe type | Skin |
| TEMPERATURE PARAMETERS | Temperature probe type | Reusable |
| TEMPERATURE PARAMETERS | Temperature probe range (atleast) | 30 - 40 degree |
| TEMPERATURE PARAMETERS | Minimum gradation of temperature probe | 1 degree Celcius |
| TEMPERATURE PARAMETERS | Accuracy in % | ± 1% |
| TEMPERATURE PARAMETERS | Duration of time of full disclosure of patient data ( in hour ) | 48 Or higher |
| TEMPERATURE PARAMETERS | Operating conditions | Capable of operating continuously in ambient temperature of 0 to 50 degree celcius and relative humidity of 15 to 90% in ideal circumstances |
| TEMPERATURE PARAMETERS | Duration of time of trending ( in hour ) | 72 Or higher |
| ALARMS | Operator can set audio visual alarm levels for low or high levels of each parameter independently | Yes |
| ALARMS | Systolic BP - low and high | Yes |
| ALARMS | Diastolic BP - low and high | Yes |
| ALARMS | Mean BP - low and high | Yes |
| ALARMS | Heart rate - low and high | Yes |
| ALARMS | SPO2 - low and high | Yes |
| ALARMS | Respiratory rate - low and high | Yes |
| ALARMS | Temperature - low and high | Yes |
| ALARMS | Duration of alarm silence in seconds | 120 |
| ALARMS | Alarm Volume should be adjustable | Yes |
| ALARMS | Activating silence of alarm | Yes |
| ALARMS | Type of alarms | Suspension of all alarms, Cuff disconnect alarm, Alarm for inflation / deflation errors, Low battery voltage indication, Sensor / wire / probe disconnection alarm |
| DISPLAY PARAMETERS | Type of display | LCD / LED / TFT |
| DISPLAY PARAMETERS | Display Size,inch with ±5% tolerance | 12, 15, 17 Or higher |
| DISPLAY PARAMETERS | User Interface of display | Rotary knob, Touchscreen, Rotary knob and Touchscreen Or higher |
| DISPLAY PARAMETERS | Inbuilt thermal Recorder (Printer) | No |
| DISPLAY PARAMETERS | Display must allow easy viewing at all ambient light levels | Yes |
| NETWORKING PARAMETERS | Ethernet port | Yes |
| NETWORKING PARAMETERS | Wireless | No |
| NETWORKING PARAMETERS | USB port | No |
| NETWORKING PARAMETERS | Connectivity provision with Central Nursing Station | Yes |
| STANDARD ACCESSORIES | Number of adult probe provided | 1 |
| STANDARD ACCESSORIES | Number of paediatric probe provided | Not provided |
| STANDARD ACCESSORIES | Number of neonatal probe provided | Not provided |
| STANDARD ACCESSORIES | Number of ECG cable provided | 1 |
| STANDARD ACCESSORIES | Number of Skin temperature probe provided | 1 |
| STANDARD ACCESSORIES | Number of Rectal temperature probe provided | Not provided |
| STANDARD ACCESSORIES | Number of Esophageal temperature probe provided | Not provided |
| STANDARD ACCESSORIES | Accessories to wall mount the monitor | Yes |
| STANDARD ACCESSORIES | Provided with protective case for clean storage and safe transport | No |
| STANDARD ACCESSORIES | Cuff provided | Adult |
| STANDARD ACCESSORIES | Cuff size provided | Medium |
| STANDARD ACCESSORIES | Number of Adult NIBP cuffs provided | 1 |
| STANDARD ACCESSORIES | Number of Paediatric NIBP cuffs provided | Not provided |
| STANDARD ACCESSORIES | Size of Disposable Neonate NIBP cuffs provided (in cm) | Not provided |
| STANDARD ACCESSORIES | Number of Disposable Neonate NIBP cuffs provided | Not provided |
| POWER REQUIREMENTS | Power supply | Single Phase 220 - 240, 50/60 Hz |
| POWER REQUIREMENTS | Mains cable to be atleast 3m length Indian type 3 pin | Yes |
| POWER REQUIREMENTS | Backup time in minutes | >=60 |
| CERTIFICATIONS | Compliance to Drugs and Cosmetic Act 1940 and Rules including Medical Device Rules (MDR) 2017 as amended till date | Yes |
| CERTIFICATIONS | Availability of valid Medical Device license or the product issued from the competent authority | Yes |
| WARRANTY | Warranty (in year) | 1 Or higher |